Its is recommended that NSE-MV filter are installed parallel in the plant room immediately prior to the vacuum reservoir. The facilities maintenance of one unit without interrupting the vacuum supply. Filters should be installed vertically with sterilisable flash at the bottom and with the strict observance of direction of flow (direction arrow marked on filter). Flow through the element is outside to in passing first through the coarse outer pre filter and then through micro fibre media. Liquid which pass down the vacuum line are separated by the filter element and collected in the drain flask. The drain flask should be inspected daily, and if any liquid is present, the flask must be valved off cleaned and autoclaved before replacement. Filter element must be replaced every six months or when the indicator show a high differential pressure (Red Zone) due to loss of vacuum. Replaced filter element must be treated as bio-hazard and disposed off accordingly.
Technical Specification :
Technical Specification :
Filter Code No. | Pipe Connection | Free Air Capacity at Atmospheric | Rarified Air capacity at 500 mm Hg vacuum | Element Code No. | ||
NL/min | Scfm | L/min | cfm | |||
MV 25 G | G 1/2 | 228 | 8 | 684 | 24 | F 1140S |
MV 30 G | G 1 | 314 | 11 | 942 | 33 | F 2010S |
MV 40 G | G 1 1/2 | 572 | 20 | 1716 | 60 | F 2020S |
MV 45 G | G 1 1/2 | 1260 | 44 | 3780 | 132 | F 2030S |
MV 50 G | G 2 | 1920 | 67 | 5760 | 201 | F 2050S |
MV 55 G | G 2 | 2780 | 97 | 8340 | 291 | F 3050S |
MV 60 G | G3 | 4250 | 149 | 12750 | 447 | F 3075S |
Larger Size are available on request
For flow of other rarefied conditions than stated above, apply the correction factor shown below to free air capacity.
Degree of Vacuum | Mm Hg | 0 | 380 | 508 | 572 | 610 | 635 | 648 | 660 | 673 | 686 |
Ins Hg | 0 | 15 | 20 | 22.5 | 24 | 25 | 25.5 | 26 | 26.5 | 27 | |
Correction Factor | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 |